Set up your experiment ====================== .. warning:: **Collision danger**: Only perform this operation together with the beamline staff. You must always pay attention to the position of endstation, detectors and sample, while performing the alignment. **Always move small steps when endstation and detector are close to each other!** .. figure:: /img/beats_endstation1.jpg :align: center :alt: BEATS experimental station *Figure 1: BEATS experimental station. Two laser lines are used to pre-align sample and detector on the beam.* Preliminary steps ----------------- #. Mount your sample on top of the tomography endstation (Figure 1 RIGHT) #. Turn on the alignment lasers #. Use the laptop close to the endstation to: #. Pre-align sample on the intersection of the laser planes (you can verify this also moving the ROT stage) #. Pre-align the detector scintillator on the same line #. Set the distance between sample and detector to the desired value Sample mount ------------ * Samples can be mounted on the tomography rotation stage with M4 screws as shown Figure 2. * A set of standard kinematic mounts from Newport is also available: `M-BK-1A `_ (download -> `drawing `_). .. figure:: /img/beats_sample_mount.png :align: center :alt: Sample mounting at BEATS *Figure 2: (LEFT) Detail of sample tomography stage. (RIGHT) The sample plate has 9 M4 holes that can be used for custom sample support.* Sample alignment ---------------- | In order to align the sample on the center of rotation of the rotary stage 4 motorized axis are needed: | | • **TOMO_Y** (vertical motion) | • **TOMO_X** (horizontal motion perpendicular to the beam) | • **SAMPLE_SX** (horizontal motion above the rotary stage) | • **SAMPLE_SZ** (horizontal motion normal to "sample top X" above the rotary stage) | The motion range of the tomography endstation (Figure 1 RIGHT) is: | | • **TOMO_X**: 60.0 mm | • **TOMO_Y**: 47.0 mm Sample alignment procedure ~~~~~~~~~~~~~~~~~~~~~~~~~~ Load the sample on the kinematic mount (for automatic alignemt of the endstation with :doc:`tomoalign` use the tungsten wire available at the beamline as sample) then: #. Perform the :doc:`hutch_search` #. Open the shutters using the :doc:`vacuum` #. Use the :ref:`Experimental GUI` to move the sample up/down until the sample is in the field of view of detector. Experimental GUI ---------------- .. figure:: /img/exp_gui.png :align: center :alt: BEATS experimental GUI *Figure 3: BEATS experimental GUI.*